Mueller-Hinton pre-poured media

| Bacteria

Results are available at https://www.clinicalmicrobiologyandinfection.com/article/S1198-743X(22)00289-0/fulltext.

Results: The best performance was observed for MHA from Becton Dickinson (BBL MHA II), bioMérieux (MHE agar) and Hardy Diagnostics, for which >97% of zone diameters were within QC ranges and >60% on target ±1 mm. The poorest performance was seen for plates from HiMedia (MHA and MHA No. 2), where 20 and 18% of readings were outside the QC ranges, respectively.